Vertical Large Capacity Shaking Incubator. Studio is processed by quality stainless steel mirror panel, easy to clean.
Vertical Large Capacity Shaking Incubator
Large Capacity Shaking Incubator
Product Feature:
♦ Double paddle, large work space design;
♦ Studio is processed by quality stainless steel mirror panel, easy to clean;
♦ Double doors hollow glass door, easy to observe the internal situation; the side is with a test hole to facilitate to test sample and to meet the demand for oxygen;
♦ Equipped with lights and UV germicidal lamp, easy to observe and sterilization
♦Imported CFC-free refrigeration compressors, refrigeration unique work designed to ensure that the compressor does not overheat due to their own self-protection shut down, so as to achieve high-quality temperature effect;
♦ Centrifugal circulating air duct, forced to work indoor air flow, the temperature more uniform;
♦ High pressure brushless DC motor control scheme, the number of revolutions measurement precision, long life, maintenance-free; available for reversing the work setting;
♦ Having a gate function, automatic door stops;
♦ Multidimensional driven, smooth operation and long service life;
♦ LCD large-screen LCD display and a color touch screen program temperature control system in the form of two control options to meet your needs;
♦ With a timer function, time setting range 0 to 9999 (minutes) hours;
♦ With power and memory function, without any interruption, crash cause data loss;
♦ With over-temperature alarm and automatic power-off function;
♦Has a self-diagnostic function;
Large Capacity Shaking Incubator
Main Technical Parameters:
Model |
LFZ-TSI-500 |
LFZ-TSI-500R |
LFZ-TSI-500A |
LFZ-TSI-500AR |
Display |
LCD |
|||
Sharking frequency |
10-300rpm |
|||
Sharking accuracy |
±1rpm |
|||
Shake plate amplitude |
Ф26mm |
Ф35mm |
||
Temp. range |
RT+5 ~65℃ |
4~65℃ |
RT+5 ~65℃ |
4~65℃ |
Temp. adjust accuracy |
±0.1℃ |
|||
Temperature uniformity |
±1 (at 37℃) |
|||
Temperature fluctuation |
±0.3℃(at 37℃) |
|||
control method |
P.I.D Microcomputer intelligent control |
|||
Size / number of plate |
830×500mm/2pcs |
|||
Maximum capacity of single layer |
100m×77or250ml×45or500ml×37or1000ml×18or2000ml×13 |
|||
Single standard capacity |
100m×45or250ml×35or500ml×24or1000ml×12or2000ml×8 |
|||
standard configuration |
2 layers(Lower fixing clamp 1000ml×12,Upper fixed clamp 500ml×24) |
|||
Timing range |
|
|||
Power (kw) |
1.8 |
2.1 |
1.9 |
2.2 |
power supply |
AC220±10% 50~60Hz |
|||
Dimensions (W*D*H) |
1060*800*1470 |
|||
Packing size (W*D*H) |
1190*895*1640 |
|||
NW |
280 |
300 |
420 |
440 |
Note:
1, Performance parameter testing under no-load conditions: ambient temperature 20° c, humidity 50%RH
2: Working ambient temperature: 5-25℃
3, In order to give full play to the best use of the