-40°C to +200°C High Low Temperature Hermetic refrigerating and heating circulators Pump, suitable for glass reactor

HRHC Series high and low temperature integrated circulating chiller heater is also called the high and low temperature circulation pump. The heating and cooling integrated machine is a high and low temperature circulation integrated machine, which is specially used for the temperature rise and fall of the reactor.

PRODUCT MODEL: HRHC- Series Category: Tag:

heating circulators Pump

HRHC Series high and low temperature integrated circulating chiller heater is also called the high and low temperature circulation pump. The heating and cooling integrated machine is a high and low temperature circulation integrated machine, which is specially used for the temperature rise and fall of the reactor.

The heat-conducting medium of the entire circulation system is a fully-closed cycle. The circulation system has a liquid expansion container. The expansion container and the liquid circulation are independent systems. The expansion container does not participate in the liquid circulation, but is mechanically connected. Even in the case of high temperature operation, the heat conducting medium in the expansion vessel will not be higher than 50 degrees.

The entire liquid cycle is a closed system. There is no absorption of water vapor at low temperatures and no oil mist at high temperatures, which increases the service life of the heat transfer oil. The same equipment and the same heat transfer medium can achieve continuous temperature control from -95 degrees to 350 degrees. The range of cooling capacity is 1KW ~ 100KW.

Common specifications:

Item

Specifications

Refrigeration

hermetic compressors

specification

enclosed non-leakage pump

In & out of nozzle

4 # (low in and high out)

Temp. control method

Microcomputer control, LCD

safety protection

delay, leakage, over current, over heat

Body material

Plastic spray anticorrosion


Suitable for 2~10L jacket reactor

Model

HRHC-10-20

HRHC-10-35

HRHC-10-40

Suit jacket reactor

2-10L

2-10L

2-10L

Volume of tank (L)

6

6

6

Heating volume (L)

2

2

2

Heating tank size (mm)

¢90×270

¢90×270

¢90×270

Oil tank size (mm)

250×100×250

250×100×250

250×100×250

Max temp. range (℃)

-20 ~ +200

-35 ~ +200

-40 ~ +200

Ambient temp. (℃)

≤25

≤25

≤25

Ambient humidity (℃)

≤60

≤60

≤60

No-load min temp.

-22℃

-32℃

-32℃

power supply

380V±10% 50HZ

380V±10% 50HZ

380V±10% 50HZ

Heating power (W)

3000

3000

3000

Cooling power (W)

2650-810

1900-1000

1900-1000

Temp. control range

-100℃-400℃

-100℃-400℃

-100℃-400℃

Temp. control accu. (℃)

±0.1

±0.1

±0.1

sensor

PT100

PT100

PT100

Medium viscosity

500C.S.T

500C.S.T

500C.S.T

Circulating pump power (w)

100

100

100

Head (m)

7

7

7

Flow (L / min)

18

18

18

Pressure (Mpa)

4

4

4

Refrigerant

R22

R22

R22

Display temp. resolution

0.1 ℃

0.1 ℃

0.1 ℃

Overall size (mm)

720 * 650 * 1120

630*550*1060

630*550*1060

Weight (KG)

85

72

72

Connector (pc)

1

1

1

Instruction Manual (pc)

1

1

1

Inlet & outlet gasket (set)

1

1

1

Stainless steel hose (pcs)

2

2

2


Suitable for 2~20L jacket reactor

Model

HRHC-20-20

HRHC-20-30

HRHC-20-40

Suit jacket reactor

2-20L

2-10L

2-20L

Volume of tank (L)

6

6

6

Heating volume (L)

2

2

2

Heating tank size (mm)

¢90×270

¢90×270

¢90×270

Oil tank size (mm)

250×100×250

250×100×250

250×100×250

Max temp. range (℃)

-20 ~ +200

-30 ~ +200

-40 ~ +200

Ambient temp. (℃)

≤25

≤25

≤25

Ambient humidity (℃)

≤60

≤60

≤60

No-load min temp.

-20℃

-32℃

-40℃

power supply

380V±10% 50HZ

380V±10% 50HZ

380V±10% 50HZ

Heating power (W)

3000

3000

3000

Cooling power (W)

3100-900

2540-1350

3720-980

Temp. control range

-100℃-400℃

-100℃-400℃

-100℃-400℃

Temp. control accu. (℃)

±0.1

±0.1

±0.1

sensor

PT100

PT100

PT100

Medium viscosity

500C.S.T

500C.S.T

500C.S.T

Circulating pump power (w)

100

100

100

Head (m)

7

7

7

Flow (L / min)

20

18

20

Pressure (Mpa)

4

4

4

Refrigerant

R22

R22

R22

In & out of nozzle

4 #

4 #

4 #

Display temp. resolution

0.1 ℃

0.1 ℃

0.1 ℃

Overall size (mm)

630*550*1060

630*550*1060

630*550*1060

Weight (KG)

68

75

84

Connector (pc)

1

1

1

Instruction Manual (pc)

1

1

1

Inlet & outlet gasket (set)

1

1

1

Stainless steel hose (pcs)

2

2

2


Suitable for 30L jacket reactor

Model

HRHC-30-20

HRHC-30-30

HRHC-30-40

Suit jacket reactor

30L

30L

30L

Volume of tank (L)

11

11

11

Heating volume (L)

2

2

2

Heating tank size (mm)

¢90×300

¢90×300

¢90×300

Oil tank size (mm)

300*300*120

300*300*120

300*300*120

Max temp. range (℃)

-20 ~ +200

-35 ~ +200

-40 ~ +200

Ambient temp. (℃)

≤25

≤25

≤25

Ambient humidity (℃)

≤60

≤60

≤60

No-load min temp.

-22℃

-35℃

-42℃

power supply

380V±10% 50HZ

380V±10% 50HZ

380V±10% 50HZ

Heating power (W)

6000

6000

6000

Cooling power (W)

6160-1820

3720-1800

4740-1140

Temp. control range

-100℃-400℃

-100℃-400℃

-100℃-400℃

Temp. control accu. (℃)

±0.1

±0.1

±0.1

sensor

PT100

PT100

PT100

Medium viscosity

500C.S.T

500C.S.T

500C.S.T

Circulating pump power (w)

280

280

280

Head (m)

10

10

10

Flow (L / min)

35

35

35

Pressure (Mpa)

4

4

4

Refrigerant

R22

R22

R22

In & out of nozzle

4 #

4 #

4 #

Display temp. resolution

0.1 ℃

0.1 ℃

0.1 ℃

Overall size (mm)

720*650*1120

720*650*1120

720*650*1120

Weight (KG)

95

107

114

Connector (pc)

1

1

1

Instruction Manual (pc)

1

1

1

Inlet & outlet gasket (set)

1

1

1

Stainless steel hose (pcs)

2

2

2


Suitable for 50L jacket reactor

Model

HRHC-50-20

HRHC-50-30

HRHC-50-40

Suit jacket reactor

50L

50L

50L

Volume of tank (L)

11

11

11

Heating volume (L)

2

2

2

Heating tank size (mm)

¢90×300

¢90×300

¢90×300

Oil tank size (mm)

300*300*120

300*300*120

300*300*120

Max temp. range (℃)

-20 ~ +200

-30 ~ +200

-40 ~ +200

Ambient temp. (℃)

≤25

≤25

≤25

Ambient humidity (℃)

≤60

≤60

≤60

No-load min temp.

-22℃

-32℃

-42℃

power supply

380V±10% 50HZ

380V±10% 50HZ

380V±10% 50HZ

Heating power (W)

6000

6000

6000

Cooling power (W)

6160-1820

3720-1800

10280-2560

Temp. control range

-100℃-400℃

-100℃-400℃

-100℃-400℃

Temp. control accu. (℃)

±0.1

±0.1

±0.1

sensor

PT100

PT100

PT100

Medium viscosity

500C.S.T

500C.S.T

500C.S.T

Circulating pump power (w)

280

280

280

Head (m)

10

10

10

Flow (L / min)

35

35

35

Pressure (Mpa)

4

4

4

Refrigerant

R22

R22

R22

In & out of nozzle

4 #

4 #

4 #

Display temp. resolution

0.1 ℃

0.1 ℃

0.1 ℃

Overall size (mm)

720*650*1120

720*650*1120

720*650*1120

Weight (KG)

95

107

124

Connector (pc)

1

1

1

Instruction Manual (pc)

1

1

1

Inlet & outlet gasket (set)

1

1

1

Stainless steel hose (pcs)

2

2

2


Suitable for 80~100L jacket reactor

Model

HRHC-100-20

HRHC-100-30

HRHC-100-40

Suit jacket reactor

80L-100L

80L-100L

80L-100L

Volume of tank (L)

30

30

30

Heating volume (L)

3

3

3

Heating tank size (mm)

¢90×300

¢90×300

¢90×480

Oil tank size (mm)

300*300*120

300*300*120

450*150*450

Max temp. range (℃)

-20 ~ +200

-30 ~ +200

-40 ~ +200

Ambient temp. (℃)

≤25

≤25

≤25

Ambient humidity (℃)

≤60

≤60

≤60

No-load min temp.

-20℃

-32℃

-42℃

power supply

380V±10% 50HZ

380V±10% 50HZ

380V±10% 50HZ

Heating power (W)

12000

12000

12000

Cooling power (W)

15700-3690

15700-3690

10280-2565

Temp. control range

-100℃-400℃

-100℃-400℃

-100℃-400℃

Temp. control accu. (℃)

±0.1

±0.1

±0.1

sensor

PT100

PT100

PT100

Medium viscosity

500C.S.T

500C.S.T

500C.S.T

Circulating pump power (w)

280

280

280

Head (m)

10


Go to Top